Z-2611-2016 Class II Terminated

Recalled by Cook Inc. — Bloomington, IN

FDA device recall Z-2611-2016 was initiated by Cook Inc. on Apr 15, 2016 and is designated Class II. Reason for recall: Increase in reports of polymer degradation of the catheter tip, resulting in tip fracture and/or separation. degradation could include loss of device function, separation of a device segment leading… The recall status is terminated (terminated Aug 4, 2017). Affected quantity: 532,395.

Recall Details

Product Type
Devices
Report Date
Aug 31, 2016
Initiation Date
Apr 15, 2016
Termination Date
Aug 4, 2017
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
532,395

Product Description

Beacon Tip Royal Flush Plus High-Flow Catheter Catheter, Intravascular, Diagnostic The catheters are intended for use in angiographic procedures by physicians trained and experienced in angiographic techniques. Standard techniques for placement of vascular access sheaths, angiographic catheters and wire guides should be employed.

Reason for Recall

Increase in reports of polymer degradation of the catheter tip, resulting in tip fracture and/or separation. degradation could include loss of device function, separation of a device segment leading to medical intervention, or complications resulting from a separated segment. Such complications include device fragments in the vascular system, genitourinary system, or other soft tissues. Fragments within the vascular system could result in embolization to the heart or lungs, or occluding blood flow to end organs.

Distribution Pattern

Worldwide Distribution - US (Nationwide) and Internationally to Algeria, Andorra, Argentina, Armenia, Australia, Austria, Bahrain, Bangladesh, Barbados, Belgium, Bermuda, Brazil, Brunei Darussalam, Bulgaria, Canada, Canary Islands, Chile, China, Colombia, Costa Rica, Cyprus, Czech Republic, Denmark, Djibouti, Dominican Republic, Ecuador, Egypt, El Salvador, Finland, France, French Polynesia, Georgia, Germany, Greece, Guadeloupe, Guatemala, Guernsey, Honduras, Hong Kong, Hungary, Iceland, India, Indonesia, Iran, Iraq, Ireland, Israel, Italy, Jamaica, Japan, Jersey, Jordan, Kazakhstan, Kenya, Korea, Kosovo, Kuwait, Latvia, Lebanon, Libyan Arab Jamahiriya, Lithuania, Luxembourg, Macau, Malaysia, Malta, Martinique, Mauritius, Mexico, Monaco, Mongolia, Morocco, Myanmar, Netherlands, Netherlands Antilles, New Caledonia, New Zealand, Nicaragua, Norway, Oman, Pakistan, "Palestinian Territory, Occupied", Panama, Paraguay, Peru, Philippines, Poland, Portugal, Puerto Rico, Qatar, Reunion, Romania, Russian Federation, Saudi Arabia, Serbia, Singapore, Slovakia, Slovenia, South Africa, Spain, Sri Lanka, Sweden, Switzerland, Taiwan, Thailand, Trinidad And Tobago, Tunisia, Turkey, Ukraine, United Arab Emirates, United Kingdom, Uruguay, Venezuela, Vietnam, U.S. Virgin Islands, Yugoslavia.

Code Information

All Lots. Catalog prefixes HNR5.0 Catalog/product number; Global/order number HNR5.0-35-100-P-10S-0 G09697 HNR5.0-35-100-P-10S-CFP G11264 HNR5.0-35-100-P-10S-PIG G09059 HNR5.0-35-100-P-8S-VCF G31403 HNR5.0-35-100-P-NS-0 G10313 HNR5.0-35-100-ST-10S-0 G09565 HNR5.0-35-100-ST-10S-PIG G09525 HNR5.0-35-110-P-10S-0 G11241 HNR5.0-35-110-P-10S-PIG G11265 HNR5.0-35-65-P-8S-VCF G12129 HNR5.0-35-70-P-10S-0 G09411 HNR5.0-35-70-P-10S-CFP G11262 HNR5.0-35-70-P-10S-PIG G08832 HNR5.0-35-70-P-10S-PIG-RICKETTS-081983 G11280 HNR5.0-35-70-P-4S-PED G11281 HNR5.0-35-70-P-NS-0 G11959 HNR5.0-35-70-ST-10S-0 G09564 HNR5.0-35-70-ST-10S-PIG G09050 HNR5.0-35-80-P-8S-NEFF-CANADA-082098 G11860 HNR5.0-35-90-P-10S-0 G09412 HNR5.0-35-90-P-10S-CFP G11266 HNR5.0-35-90-P-10S-PIG G09408 HNR5.0-35-90-P-10S-PIG-RICKETTS-081983 G11282 HNR5.0-35-90-P-6S-PED G11729 HNR5.0-35-90-P-8S-VCF G12799 HNR5.0-35-90-ST-10S-0 G11641 HNR5.0-35-90-ST-10S-PIG G09197 HNR5.0-38-100-P-10S-0 G11267 HNR5.0-38-100-P-10S-CFP G11268 HNR5.0-38-100-P-10S-PIG G09285 HNR5.0-38-100-P-NS-0 G27874 HNR5.0-38-110-P-10S-0 G09461 HNR5.0-38-110-P-10S-PIG G11269 HNR5.0-38-70-P-10S-0 G11270 HNR5.0-38-70-P-10S-CFP G11271 HNR5.0-38-70-P-10S-PIG G11272 HNR5.0-38-70-P-NS-0 G11289 HNR5.0-38-80-P-10S-OPC G27877 HNR5.0-38-90-P-10S-0 G11273 HNR5.0-38-90-P-10S-CFP G11274 HNR5.0-38-90-P-10S-PIG G11275

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-2611-2016
openFDA file date
Oct 23, 2026
Loaded by RegDataLab
Oct 26, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.