Z-2618-2014 Class II Terminated

Recalled by Becton Dickinson & Co. — Sparks, MD

FDA device recall Z-2618-2014 was initiated by Becton Dickinson & Co. on Jul 23, 2014 and is designated Class II. Reason for recall: The performance of the assay no longer conforms to this statement in the Analytical Specificity section of the Package Insert: One hundred-eleven (111) out of 111 MSSA strains tested at extremely… The recall status is terminated (terminated May 9, 2016). Affected quantity: 585,048 tests.

Recall Details

Product Type
Devices
Report Date
Sep 17, 2014
Initiation Date
Jul 23, 2014
Termination Date
May 9, 2016
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
585,048 tests

Product Description

The BD MAX MRSA Assay, catalogue #442953. An automated qualitative in vitro diagnostic test for the direct detection of Methicillin-resistant Staphylococcus aureus (MRSA) DNA from nasal swabs in patie.

Reason for Recall

The performance of the assay no longer conforms to this statement in the Analytical Specificity section of the Package Insert: One hundred-eleven (111) out of 111 MSSA strains tested at extremely high concentrations (> 106 CFU/swab), produced negative results with the BD MAX MRSA Assay. One of the samples tested with each of the CAP surveys MRS5-A 2014 and MRS5-B 2014 yielded false positive results.

Distribution Pattern

US Distribution including the states of : CA, FL, HI, IL, IN, MD, MI, MN, MT, NC, ND, NJ, NM, NV, NY, OH, OK, OR, PA, SC, TN, TX, UT, VA, WA and WI.

Code Information

All Lots with the expiration date January 21, 2014 or later

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-2618-2014
openFDA file date
Oct 23, 2026
Loaded by RegDataLab
Oct 25, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.