Z-2635-2023 Class II Ongoing

Recalled by Fresenius Kabi USA, LLC — North Andover, MA

FDA device recall Z-2635-2023 was initiated by Fresenius Kabi USA, LLC on Apr 22, 2022 and is designated Class II. Reason for recall: Start-up time for the flow rate range greater than 50 ml/hr, and less than 200 ml/hr may exceed the specified 6 minutes, depending on the characteristics of air bubble entrapment within the… The recall status is ongoing. Affected quantity: 1344 pumps.

Recall Details

Product Type
Devices
Report Date
Oct 4, 2023
Initiation Date
Apr 22, 2022
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
1344 pumps

Product Description

Ivenix Infusion System (IIS), Large Volume Pump (LVP) Software: 5.0.0 and 5.0.1 Part Number: LVP-SW-0004. Indicated for use in a hospital and in outpatient care environments for the controlled administration of fluids through clinically accepted routes of administration.

Reason for Recall

Start-up time for the flow rate range greater than 50 ml/hr, and less than 200 ml/hr may exceed the specified 6 minutes, depending on the characteristics of air bubble entrapment within the Intermediate Pumping Chamber (IPC) of the administration set

Distribution Pattern

US Nationwide distribution in the states of NJ, WI.

Code Information

(UDI): 00811505030122 Software Version: 5.0.0 and 5.0.1