Z-2635-2023 Class II Ongoing

Recalled by Fresenius Kabi USA, LLC — North Andover, MA

Recall Details

Product Type
Devices
Report Date
October 4, 2023
Initiation Date
April 22, 2022
Termination Date
N/A
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
1344 pumps

Product Description

Ivenix Infusion System (IIS), Large Volume Pump (LVP) Software: 5.0.0 and 5.0.1 Part Number: LVP-SW-0004. Indicated for use in a hospital and in outpatient care environments for the controlled administration of fluids through clinically accepted routes of administration.

Reason for Recall

Start-up time for the flow rate range greater than 50 ml/hr, and less than 200 ml/hr may exceed the specified 6 minutes, depending on the characteristics of air bubble entrapment within the Intermediate Pumping Chamber (IPC) of the administration set

Distribution Pattern

US Nationwide distribution in the states of NJ, WI.

Code Information

(UDI): 00811505030122 Software Version: 5.0.0 and 5.0.1