Z-2645-2016 Class II Terminated

Recalled by Medtronic Neuromodulation — Minneapolis, MN

FDA device recall Z-2645-2016 was initiated by Medtronic Neuromodulation on Jun 23, 2016 and is designated Class II. Reason for recall: Due to the potential for pocket adaptor conductor wire fractures involving DBS Pocket Adaptor Model 64001 and Model 64002, a safety notification was sent to healthcare professionals to reinforce… The recall status is terminated (terminated Sep 21, 2017). Affected quantity: 7,542 in the US and 12,999 outside the US.

Recall Details

Product Type
Devices
Report Date
Aug 31, 2016
Initiation Date
Jun 23, 2016
Termination Date
Sep 21, 2017
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
7,542 in the US and 12,999 outside the US.

Product Description

DBS Pocket Adaptor, Model 64001 and Model 64002. The Medtronic Models 64001 (1x4) and 64002 (2x4) Pocket Adaptors can be used as a part of a neurostimulation system for deep brain stimulation. The pocket adaptor is intended to be implanted with the new replacement neurostimulator in the same pocket used for the explanted neurostimulator. Implanting in the same neurostimulator pocket allows for a single-incision procedure.

Reason for Recall

Due to the potential for pocket adaptor conductor wire fractures involving DBS Pocket Adaptor Model 64001 and Model 64002, a safety notification was sent to healthcare professionals to reinforce device specific labeling for the handling of pocket adaptors and system integrity checking during implant procedures.

Distribution Pattern

Worldwide Distribution.

Code Information

Notification is not lot specific.

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-2645-2016
openFDA file date
Oct 23, 2026
Loaded by RegDataLab
Oct 26, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.