Z-2658-2023 Class II Ongoing

Recalled by Thoratec Corp. — Pleasanton, CA

FDA device recall Z-2658-2023 was initiated by Thoratec Corp. on Aug 7, 2023 and is designated Class II. Reason for recall: Inability to start and/or complete the coring procedure due to inadequate sharpness of the Apical Coring Knife, provided with Left Ventricular Assist System, could result in an extended procedure or… The recall status is ongoing. Affected quantity: 2,681.

Recall details

Product Type
Devices
Report Date
Oct 11, 2023
Initiation Date
Aug 7, 2023
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
2,681

Product description

Heartmate 3 LVAS Implant Kit, REF: 106524US, 106524INT; Heartmate XVE & HeartMate II Apical Coring Knife, REF: 1050

Reason for recall

Inability to start and/or complete the coring procedure due to inadequate sharpness of the Apical Coring Knife, provided with Left Ventricular Assist System, could result in an extended procedure or surgical time while the backup coring knife is acquired.

Distribution pattern

Worldwide - US Nationwide distribution including in the states of AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, IA, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, NC, NE, NJ, NM, NV, NY, OH, OK, OR, PA, SC, TN, TX, UT, VA, WA, WI, WV and the countries of Australia, Austria, Belgium, Brazil, Bulgaria, Canada, Chile, Colombia , Czech Republic, Denmark, Egypt, Estonia, Finland, France, Germany, Greece, Hong Kong, India, Iran, Iraq, Ireland, Israel, Italy, Japan, Korea, Kuwait, Lebanon, Lithuania, Macedonia, Malaysia, Mexico, Morocco, Netherlands, New Zealand, Norway, Poland, Portugal, Qatar, Romania, Russia, Saudi Arabia, Serbia, Singapore, Slovenia, South Africa, Spain, Sweden, Switzerland, Taiwan, Thailand, Tunisia, Turkey, United Arab Emirates, United Kingdom.

Code information

REF/UDI-DI/Serial Numbers: 106524US/00813024013297/MLP-034137 to MLP-034744, MLP-035485 to MLP-035714, MLP-036330 to MLP-037708, MLP-038104 to MLP-038999, MLP-039000 to MLP-039999, MLP-040000 to MLP-040999, MLP-041000 to MLP-041621; 106524INT/00813024011712 (OUS)/MLP-032482, MLP-035542 to MLP-035667, MLP-037218 to MLP-037527, MLP-038008 to MLP-038989, MLP-039014 to MLP-039945, MLP-040009 to MLP-040896, MLP-041013 to MLP-041589; 1050/00813024010227: 135021-135025, 134241-134246, 134248, 134250, 134251, 134240, 135060, 135061, 135062, 134247, 132849-132853, 134239; Apical Coring Knife Distributed Separately Serial Number Range: 132849 to 135062

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-2658-2023
openFDA file date
Nov 19, 2026
Loaded by RegDataLab
Nov 20, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Need the full FDA record for Thoratec Corp.? Ask Claude or ChatGPT with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. The API key is free: no card.

Get a free API key

How the connector works

Source: openFDA record for Z-2658-2023. Report a data issue: hello@regdatalab.com.