Z-2291-2025 Class I Ongoing

Recalled by Thoratec LLC — Pleasanton, CA

FDA device recall Z-2291-2025 was initiated by Thoratec LLC on Jun 23, 2025 and is designated Class I. Reason for recall: Due to misaligned V-Lock feature on the Mobile Power Unit power cords, which prevents the locking mechanism from properly engaging with the Mobile Power Unit and may cause unintended disconnect. The recall status is ongoing. Affected quantity: 604.

Recall details

Product Type
Devices
Report Date
Aug 20, 2025
Initiation Date
Jun 23, 2025
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
604

Product description

Brand Name: HeartMate Product Name: Mobile Power Unit Model/Catalog Number: 107754 Software Version: N/A Product Description: The HeartMate 3" Left Ventricular Assist System (LVAS) is a set of equipment and materials that together comprise a medical device designed to provide therapeutic benefit to those affected with advanced heart failure. In service, the LVAS assumes some or all of the workload of the left ventricle, thereby restoring the patient's systemic perfusion while palliating the underlying pathology. The LVAS features a Left Ventricular Assist Device (LVAD), a blood pump intended for long-term implantation in such patients, an extracorporeal Controller, plus all of the features, controls, attachments, interfaces, power sources, supporting equipment, labeling, and tools required to achieve the desired therapeutic benefit. The HeartMate II Left Ventricular Assist System is an axial-flow, rotary ventricular assist system that generates flows up to 10 liters per minute (lpm). One end of the Left Ventricular Assist Device is attached to the apex of the left ventricle; the other end of the device connects to the ascending aorta. The Left Ventricular Assist Device diverts blood from the weakened left ventricle and propels it to the aorta. A small external computer, the System Controller, monitors system operation. A driveline, which passes through the patient s abdomen, connects the implanted pump to the System Controller. The system is powered by a Power Module or Mobile Power Unit (MPU) that is connected to an AC electrical outlet, or by two HeartMate 14 Volt Lithium-Ion batteries. The LVAS may be used in any of two configurations. First, line power may be utilized through the Power Module or the Mobile Power Unit (MPU) to run the LVAD indefinitely, convenient for sedentary or sleeping periods. Second, portable Battery power may be utilized for limited periods, convenient for active periods. The MPU AC power cord is distributed either as part of the complete MPU device package or as a standalone accessory. Component: Yes, the MPU is a component of the HeartMate 3 and HeartMate II Left Ventricular Assist Systems.

Reason for recall

Due to misaligned V-Lock feature on the Mobile Power Unit power cords, which prevents the locking mechanism from properly engaging with the Mobile Power Unit and may cause unintended disconnect.

Distribution pattern

Worldwide - US Nationwide distribution including in the states of AL, AR, AZ, CA, CO, CT, DE, FL, GA, IA, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, NC, NE, NJ, NM, NY, OH, OK, PA, SC, TN, TX, UT, VA, WA, WI, WV and the countries of Canada, Japan.

Code information

Lot Code: US Model No 107754; UDI-DI 05415067038234 Serial Numbers, see Attachment F.

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-2291-2025
openFDA file date
Nov 19, 2026
Loaded by RegDataLab
Nov 20, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Need the full FDA record for Thoratec LLC? Ask Claude or ChatGPT with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. The API key is free: no card.

Get a free API key

How the connector works

Source: openFDA record for Z-2291-2025. Report a data issue: hello@regdatalab.com.