Z-2662-2018 Class II Terminated

Recalled by Smiths Medical ASD Inc. — Minneapolis, MN

Recall Details

Product Type
Devices
Report Date
August 15, 2018
Initiation Date
April 13, 2018
Termination Date
January 30, 2023
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
276 units

Product Description

Medfusion Syringe Infusion Pump, containing the Barrel Clamp Guide mechanism

Reason for Recall

Certain Barrel Clamp Guides may contain a ridge that could potentially lead to spring slippage, resulting in the inability of the pump to recognize a loaded syringe or the pump, misidentifying the size of a loaded syringe.

Distribution Pattern

US distribution to AL, AZ, CA, DC, FL, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, NJ, NM, NY, OH, OK, PA, TN, TX, UT, VA, WA. International distribution to Canada and Lebanon.

Code Information

Barrel Clamp Guide mechanism number: G6000716; Barrel Clamp Guide lot numbers: P0239588, P0321444, and P0462421