Z-2662-2018 Class II Terminated

Recalled by Smiths Medical ASD Inc. — Minneapolis, MN

FDA device recall Z-2662-2018 was initiated by Smiths Medical ASD Inc. on Apr 13, 2018 and is designated Class II. Reason for recall: Certain Barrel Clamp Guides may contain a ridge that could potentially lead to spring slippage, resulting in the inability of the pump to recognize a loaded syringe or the pump, misidentifying the… The recall status is terminated (terminated Jan 30, 2023). Affected quantity: 276 units.

Recall Details

Product Type
Devices
Report Date
Aug 15, 2018
Initiation Date
Apr 13, 2018
Termination Date
Jan 30, 2023
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
276 units

Product Description

Medfusion Syringe Infusion Pump, containing the Barrel Clamp Guide mechanism

Reason for Recall

Certain Barrel Clamp Guides may contain a ridge that could potentially lead to spring slippage, resulting in the inability of the pump to recognize a loaded syringe or the pump, misidentifying the size of a loaded syringe.

Distribution Pattern

US distribution to AL, AZ, CA, DC, FL, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, NJ, NM, NY, OH, OK, PA, TN, TX, UT, VA, WA. International distribution to Canada and Lebanon.

Code Information

Barrel Clamp Guide mechanism number: G6000716; Barrel Clamp Guide lot numbers: P0239588, P0321444, and P0462421

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-2662-2018
openFDA file date
Oct 23, 2026
Loaded by RegDataLab
Oct 26, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.