Z-2694-2017 Class II Terminated
Recalled by St. Jude Medical, Inc.
— Plano, TX
FDA device recall Z-2694-2017 was initiated by St. Jude Medical, Inc. on Jun 2, 2017 and is designated Class II. Reason for recall: The firm discovered a risk of loss of Neuromodulation therapy due to exposure of the IPG to monopolar electrosurgical devices (ESU) during medical procedures. The recall status is terminated (terminated Aug 24, 2022). Affected quantity: 54 units.
Recall details
- Product Type
- Devices
- Report Date
- Jul 19, 2017
- Initiation Date
- Jun 2, 2017
- Termination Date
- Aug 24, 2022
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 54 units
Product description
Infinity 5, Implantable Pulse Generator, REF 6661, Rx only, STERILE EO Product Usage: The Infinity IPG is an electronic device designed to be connected to one or two extensions. It is powered by a hermetically sealed battery within a titanium case and uses microelectronic circuitry to generate constant-current electrical stimulation. The IPG is conductive on all sides, which allows the IPG case (also called a can) to be used as an anode for monopolar stimulation. The IPG communicates wirelessly with system programmers and controllers, and IPGs are available in small and large sizes to accommodate different power needs.
Reason for recall
The firm discovered a risk of loss of Neuromodulation therapy due to exposure of the IPG to monopolar electrosurgical devices (ESU) during medical procedures.
Distribution pattern
US Nationwide Distribution
Code information
ALX120.1 ALX227.1 ALY287.1 ALY390.1 ALY471.1 ALY534.1 ALY560.1 ALY668.1 ALZ030.1 AMN762.1 AMN904.1 AMN990.1 AMP134.1 AMP169.1 AMP189.1 AMP246.1 AMS702.1 AMT135.1 AMT160.1 AMT357.1 ANG915.1 ANG951.1 ZA3446.1 ZA3511.1 ZB0872.1 ZB1144.1 ZB1181.1 ZB1188.1 ZB1190.1 ZB1200.1 ZB1213.1 ZB1218.1 ZB1225.1 ZB1422.1 ZB1463.1 ZB1507.1 ZK3289.1 ZS0594.1 ZS0699.1 ZS0772.1 ZW0246.1 ZW0268.1 ZW0273.1 ZW0283.1 ZW0364.1 ZW8776.1 ZW8796.1 ZW8863.1 ZW8879.1 ZW8887.1 ZW8904.1 ZW8965.1 ZW8972.1 ZX5593.1
Source
- Dataset
- openFDA device recall enforcement reports
- FDA record
- This recall in openFDA (JSON)
- Enforcement Report
- Search FDA's Enforcement Report for Z-2694-2017
- openFDA file date
- Nov 25, 2026
- Loaded by RegDataLab
- Nov 26, 2026
These dates say when openFDA published the data file and when RegDataLab last loaded it.