Z-1219-2023 Class II Ongoing

Recalled by St. Jude Medical — Atlanta, GA

FDA device recall Z-1219-2023 was initiated by St. Jude Medical on Feb 7, 2023 and is designated Class II. Reason for recall: Radiated Radiofrequency (RF) emissions of the CardioMEMS Patient Electronics Systems (PES) Models CM1000, CM1010 and CM1100 and CardioMEMS Hospital Electronics Systems (HES) Model CM3000 exceed… The recall status is ongoing. Affected quantity: 200 units.

Recall details

Product Type
Devices
Report Date
Mar 15, 2023
Initiation Date
Feb 7, 2023
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
200 units

Product description

CardioMEMS HF System Patient Electronics System (PES), Model number CM1010, The CardioMEMS HF System provides pulmonary artery (PA) hemodynamic data used for the monitoring and management of heart failure (HF) patients. The system measures changes in PA pressure which physicians use to initiate or modify heart failure treatment.

Reason for recall

Radiated Radiofrequency (RF) emissions of the CardioMEMS Patient Electronics Systems (PES) Models CM1000, CM1010 and CM1100 and CardioMEMS Hospital Electronics Systems (HES) Model CM3000 exceed certain emissions limits and have the potential to cause interference with other medical devices, such as implantable neurostimulators, pacemakers, cardiac defibrillators, continuous glucose monitors, or other bio-wearable sensors when in close proximity to an active Hospital or Patient Electronics System (i.e., during readings).

Distribution pattern

Worldwide distribution - US Nationwide and the countries of Argentina, Australia, Austria, Belgium, Canada, Colombia, Czech Republic, Denmark, Finland, France, Germany, Ireland, Italy, Kuwait, Netherlands, Portugal, Saudi Arabia, Spain, Sweden, Switzerland, UK, United Arab Emirates.

Code information

UDI/DI 05414734509176, Serial Numbers: M140300439, M140300440, M140300441, M140300445, M140300446, M140300447, M140300449, M140300450, M140300451, M140300454, M140300458, M140300459, M140300460, M140300463, M140500480, M140500486, M140800987, M140800988, M140800992, M140800995, M140801002, M140801003, M140801004, M140801007, M140801010, M140801011, M141001528, M141001529, M141001530, M141001531, M141001536, M141001539, M141001541, M141001544, M141001545, M141001546, M141101678, M141101681, M141101682, M141101683, M141101684, M141101685, M141101686, M141101693, M141101696, M141101697, M141101700, M141101701, M141101703, M141101705, M141101706, M150102038, M150102039, M150102049, M150102052, M150102054, M150102055, M150102059, M150102061, M150102062, M150102064, M150102066, M150102399, M150102402, M150102404, M150102409, M150102410, M150102414, M150102416, M150202537, M150202538, M150202541, M150202542, M150202544, M150202545, M150302658, M150302660, M150302662, M150302664, M150302666, M150302667, M150302669, M150302673, M150302680, M150302682, M150302683, M150302687, M150302689, M150302690, M150302692, M150302694, M150302695, M150302702, M150302703, M150302707, M150302711, M150302712, M150302713, M150302716, M150302721, M150302723, M150302724, M150302725, M150302729, M150302737, M150302739, M150302742, M150302747, M150302748, M150302751, M150302755, M150302757, M150302758, M150302759, M150302762, M150302769, M150302773, M150302777, M150302778, M150302779, M150302780, M150302782, M150302783, M150302786, M150403048, M150403050, M150403051, M150403057, M150403066, M150805139, M150805143, M150805151, M150805154, M150805155, M150805156, M150805158, M150805160, M150805161, M150805162, M150805166, M150905677, M150905680, M150905682, M150905683, M150905685, M150905688, M150905689, M150905691, M150905692, M150905697, M150905698, M150905699, M150905702, M150905703, M150905704, M150905706, M150905707, M150905708, M150905709, M150905711, M150905714, M151206628, M151206631, M151206633, M151206636, M151206637, M151206641, M151206644, M151206645, M151206647, M151206648, M151206651, M151206655, M151206659, M151206662, M151206665, M160206849, M160206853, M160206854, M160206857, M160206861, M160206867, M160206869, M160206872, M160206876, M160206879, M160206880, M160206882, M160206885, M160206888, M160206896, M160206899, M160206900, M160206902, M160206903, M160206906, M160807699, M160807700, M160807701, M160807772.

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-1219-2023
openFDA file date
Nov 25, 2026
Loaded by RegDataLab
Nov 26, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Need the full FDA record for St. Jude Medical? Ask Claude with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. Free to start, no card.

Question to ask in Claude

Use RegDataLab to give me an FDA risk summary for St. Jude Medical: recent inspections and their outcomes, warning letters, recalls and import alerts by site, with source links and dates.

Ask Claude about St. Jude Medical →

Source: openFDA record for Z-1219-2023. Report a data issue: hello@regdatalab.com.