Z-2723-2014 Class II Terminated

Recalled by Instrumed International, Inc. — Schaumburg, IL

FDA device recall Z-2723-2014 was initiated by Instrumed International, Inc. on May 30, 2014 and is designated Class II. Reason for recall: Additional caution statements have been added to the Instructions For Use for monopolar electrodes. 1) To avoid tissue carbonation, the operation voltage of the HF generator must not exceed 650 peak… The recall status is terminated (terminated Nov 29, 2017). Affected quantity: 156 instruments.

Recall Details

Product Type
Devices
Report Date
Oct 8, 2014
Initiation Date
May 30, 2014
Termination Date
Nov 29, 2017
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
156 instruments

Product Description

Reusable laparoscopic instruments are packaged as non-sterile. Cleaning and sterilization of devices must occur prior to use. The instruments are intended to be used by qualified physicians to provide access, illumination and visualization of internal structures and for manipulating soft tissue (grasping, cutting, coagulating, dissecting and suturing) in a wide variety of diagnostic and therapeutic laparoscopic/ urologic closed and minimally invasive procedures. It is inserted via a trocar sleeve and is used to prepare, coagulate and cut tissue.

Reason for Recall

Additional caution statements have been added to the Instructions For Use for monopolar electrodes. 1) To avoid tissue carbonation, the operation voltage of the HF generator must not exceed 650 peak voltage (Vp) for all monopolar electrodes. and 2) For all monopolar electrodes, the useful life of this product is less than or equal to 50 cycles and less than or equal to 2 years.

Distribution Pattern

Distribution US Nationwide in the states of: AL, AR, AZ, CA, CO, CT, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, WA, WI, WV, and WY.

Code Information

1) Instrumed Item No.: 250-08680; Boss Instruments Catalog Code: 35-8015; Lot Number: 030411, 100110, 110210; 2) Instrumed Item No.: 250-08700; Boss Instruments Catalog Code: 35-8016; Lot Number: 030111; 3) Instrumed Item No.: 250-08700; Boss Instruments Catalog Code: 35-8016; Lot Number: 30111; 4) Instrumed Item No.: 250-08760; Boss Instruments Catalog Code: 35-8020; Lot Number: 020411, 070111, 100412, 110412; 5) Instrumed Item No.: 250-08770; Boss Instruments Catalog Code: 35-8020L; Lot Number: 020112, 110211, 120310; 6) Instrumed Item No.: 250-08800; Boss Instruments Catalog Code: 35-8025; Lot Number: 060110, 100112; 7) Instrumed Item No.: 250-08820; Boss Instruments Catalog Code: 35-8026 ; Lot Number: 032811; 8) Instrumed Item No.: 250-08860; Boss Instruments Catalog Code: 35-8030; Lot Number: 050413, 070412, 110510, 090110; 9) Instrumed Item No.: 250-08880; Boss Instruments Catalog Code: 35-8030L; Lot Number: 090110; 10) Instrumed Item No.: 250-08890; Boss Instruments Catalog Code: 35-8035; Lot Number: 100213; 11) Instrumed Item No.: 250-08930; Boss Instruments Catalog Code: 35-8040; Lot Number: 010211, 020213, 020811, 021711, 040212, 040311, 080212, 100310, 110113, 120413; 12) Instrumed Item No.: 250-08960; Boss Instruments Catalog Code: 35-8040L; Lot Number: 030413, 040412, 040513, 120113; 13) Instrumed Item No.: 250-08990; Boss Instruments Catalog Code: 35-8042; Lot Number: 050312; 14) Instrumed Item No.: 250-09000; Boss Instruments Catalog Code: 35-8042L; Lot Number: 100406; 15) Instrumed Item No.: 250-09010; Boss Instruments Catalog Code: 35-8045; Lot Number: 010311, 020311, 090312, 100212, 120112, 120213

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-2723-2014
openFDA file date
Oct 27, 2026
Loaded by RegDataLab
Oct 29, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.