Z-2735-2014 Class II Terminated
Recall Details
- Product Type
- Devices
- Report Date
- October 8, 2014
- Initiation Date
- June 17, 2013
- Termination Date
- March 14, 2016
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 1,660 extraction trays
Product Description
Verigene BC-GP Kit (Catalog number 20-005-018); 1) 20 Verigene BC-GP Test Cartridges; are qualitative multiplexed in vitro diagnostic tests for simultaneous detection and identification of selected gram-positive (BC-GP) and gram-negative (BC-GN) bacteria and resistance markers.
Reason for Recall
There is a specific Extraction Tray lot containing Tips that may slightly increase the occurrence of Process SP Processing Error when used due to the Processor SP experiencing difficulty releasing the Tip to the final location in the Tip Assembly. The difficult releasing the Tip can result in a Processing Error and occasionally the Tip to be dropped into the Processor SP.
Distribution Pattern
Worldwide Distribution:US (Nationwide) including states of: AZ, CA, FL, IL, KY, MD, MN, MT, OH, PA, TN, TX, VA, WA, WI; and countries of: Belgium, China, and Saudi Arabia.
Code Information
Part Number: 20-009-018; Lot Numbers: 051713018B, 052413018C