Z-2736-2014 Class II Terminated

Recalled by Nanosphere, Inc. — Northbrook, IL

FDA device recall Z-2736-2014 was initiated by Nanosphere, Inc. on Jun 17, 2013 and is designated Class II. Reason for recall: There is a specific Extraction Tray lot containing Tips that may slightly increase the occurrence of Process SP Processing Error when used due to the Processor SP experiencing difficulty releasing the… The recall status is terminated (terminated Mar 14, 2016). Affected quantity: 20 extraction trays.

Recall details

Product Type
Devices
Report Date
Oct 8, 2014
Initiation Date
Jun 17, 2013
Termination Date
Mar 14, 2016
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
20 extraction trays

Product description

Verigene BC-GN Kit (Catalog number 20-005-021); 1) 20 Verigene BC-GN Test Cartridges; are qualitative multiplexed in vitro diagnostic tests for simultaneous detection and identification of selected gram-positive (BC-GP) and gram-negative (BC-GN) bacteria and resistance markers.

Reason for recall

There is a specific Extraction Tray lot containing Tips that may slightly increase the occurrence of Process SP Processing Error when used due to the Processor SP experiencing difficulty releasing the Tip to the final location in the Tip Assembly. The difficult releasing the Tip can result in a Processing Error and occasionally the Tip to be dropped into the Processor SP.

Distribution pattern

Worldwide Distribution:US (Nationwide) including states of: AZ, CA, FL, IL, KY, MD, MN, MT, OH, PA, TN, TX, VA, WA, WI; and countries of: Belgium, China, and Saudi Arabia.

Code information

Part Number: 20-009-021; Lot Numbers: 052113021C

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-2736-2014
openFDA file date
Nov 19, 2026
Loaded by RegDataLab
Nov 20, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Need the full FDA record for Nanosphere, Inc.? Ask Claude or ChatGPT with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. The API key is free: no card.

Get a free API key

How the connector works

Source: openFDA record for Z-2736-2014. Report a data issue: hello@regdatalab.com.