Z-2754-2018 Class II Terminated

Recalled by Degania Silicone, Ltd. — Dganya Bet, N/A

Recall Details

Product Type
Devices
Report Date
September 26, 2018
Initiation Date
June 21, 2018
Termination Date
October 25, 2024
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
153

Product Description

Temperature sensor Foley catheter 18FR, Catalogue Number 102205101880AB Product Usage: This is sterile, single use, two-way silicone Foley Catheter with a thermistor embedded in the third lumen. It is intended for use in the drainage/ collection of urine from the urinary bladder and simultaneous monitoring of the body core temperature during surgery or post-surgical period.

Reason for Recall

The catheters are suspected to contain defective sensors; if used, defective catheter will show lower body temperature than the actual body temperature of a patient.

Distribution Pattern

Worldwide Distribution - US Nationwide in the states: IL, NC, NY and the countries of : Belgium, Israel, Japan, and Mexico.

Code Information

S17005574