Z-2766-2018 Class II Terminated
Recalled by Degania Silicone, Ltd.
— Dganya Bet, N/A
FDA device recall Z-2766-2018 was initiated by Degania Silicone, Ltd. on Jun 21, 2018 and is designated Class II. Reason for recall: The catheters are suspected to contain defective sensors; if used, defective catheter will show lower body temperature than the actual body temperature of a patient. The recall status is terminated (terminated Oct 25, 2024). Affected quantity: 92150.
Recall details
- Product Type
- Devices
- Report Date
- Sep 26, 2018
- Initiation Date
- Jun 21, 2018
- Termination Date
- Oct 25, 2024
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 92150
Product description
Foley Catheter with temperature sensor 400TM 16FR, Catalogue Number 102201101680TY Product Usage: This is sterile, single use, two-way silicone Foley Catheter with a thermistor embedded in the third lumen. It is intended for use in the drainage/ collection of urine from the urinary bladder and simultaneous monitoring of the body core temperature during surgery or post-surgical period.
Reason for recall
The catheters are suspected to contain defective sensors; if used, defective catheter will show lower body temperature than the actual body temperature of a patient.
Distribution pattern
Worldwide Distribution - US Nationwide in the states: IL, NC, NY and the countries of : Belgium, Israel, Japan, and Mexico.
Code information
S17002091, S17002440, S17001607, S17002441, S17002558, S17002559, S17002561, S17002562, S17002563, S17002681, S17003391, S17004379, S17002560, S17002682, S17003456, S17003219, S17003301, S17003302, S17004772, S17004376, S17004377, S17004378, S17005117, S17005029, S17005281, S17005983, S17006720, S17004509, S17004510, S17004511, S17004771, S17004902, S17005028, S17005030, S17005529, S17005984, S17005985, S17004380, S17005979, S17006723, S17005980, S17005981, S17005982, S17005986, S17006718, S17006719, S17006721, S17006722, S17006724, S17006725, S17007360, S17007459, S17007622, S17002986, S17006363, S17007623, S17002987, S17007806, S17008017, S17002090
Source
- Dataset
- openFDA device recall enforcement reports
- FDA record
- This recall in openFDA (JSON)
- Enforcement Report
- Search FDA's Enforcement Report for Z-2766-2018
- openFDA file date
- Nov 24, 2026
- Loaded by RegDataLab
- Nov 25, 2026
These dates say when openFDA published the data file and when RegDataLab last loaded it.
Need the full FDA record for Degania Silicone, Ltd.? Ask Claude or ChatGPT with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. The API key is free: no card.
Get a free API key
How the connector works