Z-2785-2020 Class II Terminated

Recalled by Flower Orthopedics Corporation — Horsham, PA

FDA device recall Z-2785-2020 was initiated by Flower Orthopedics Corporation on Jul 8, 2020 and is designated Class II. Reason for recall: The contract manufacturer of the calcaneus plate inadvertently used the incorrect material to manufacture the plates (a titanium alloy rather than pure titanium). The recall status is terminated (terminated Oct 8, 2020). Affected quantity: 25.

Recall details

Product Type
Devices
Report Date
Aug 19, 2020
Initiation Date
Jul 8, 2020
Termination Date
Oct 8, 2020
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
25

Product description

Calcaneus Plate, Extensile, Small, Right, Catalog Number CPF 201 - Product Usage: intended to be used for internal fixation of fractures and reconstruction of bones of the rearfoot, including the calcaneus.

Reason for recall

The contract manufacturer of the calcaneus plate inadvertently used the incorrect material to manufacture the plates (a titanium alloy rather than pure titanium).

Distribution pattern

US Nationwide distribution including in the states of AZ, FL, and OH.

Code information

Lot Number 1909660034; UDI 00840118115674

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-2785-2020
openFDA file date
Nov 25, 2026
Loaded by RegDataLab
Nov 26, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Need the full FDA record for Flower Orthopedics Corporation? Ask Claude with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. Free to start, no card.

Question to ask in Claude

Use RegDataLab to give me an FDA risk summary for Flower Orthopedics Corporation: recent inspections and their outcomes, warning letters, recalls and import alerts by site, with source links and dates.

Ask Claude about Flower Orthopedics Corporation →

Source: openFDA record for Z-2785-2020. Report a data issue: hello@regdatalab.com.