Z-2821-2026 Class II Ongoing

Recalled by Instrumentation Laboratory — Bedford, MA

FDA device recall Z-2821-2026 was initiated by Instrumentation Laboratory on May 8, 2026 and is designated Class II. Reason for recall: Potential for Bar Code Reader (BCR) wands, used with the GEM Premier 5000 and GEM Premier 7000 systems, to assign patient results to an incorrect patient ID. The recall status is ongoing. Affected quantity: 122.

Recall Details

Product Type
Devices
Report Date
August 5, 2026
Initiation Date
May 8, 2026
Termination Date
N/A
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
122

Product Description

GEM Premier 5000 with IQM2 Model/Catalog Number: 00024019255 and GEM Premier 7000 with iQM3 Model/Catalog Number: 00000015279 Portable blood analyzer system. Affected Accessory: Bar Code Reader (BCR) Wand Part No. 00024015859

Reason for Recall

Potential for Bar Code Reader (BCR) wands, used with the GEM Premier 5000 and GEM Premier 7000 systems, to assign patient results to an incorrect patient ID.

Distribution Pattern

US Nationwide distribution.

Code Information

GEM Premier 5000 Part number: 00024019255, UDI: 08426950870227 GEM Premier 7000 Part number: 00000015279, UDI: 0195226000924 Affected Accessory: Bar Code Reader (BCR) Wand that received a configuration update on or after 8 April 2026. Part No. 00024015859; UDI No. 08426950869634