Z-2826-2026 Class II Ongoing

Recalled by Instrumentation Laboratory — Bedford, MA

FDA device recall Z-2826-2026 was initiated by Instrumentation Laboratory on May 8, 2026 and is designated Class II. Reason for recall: Potential for Bar Code Reader (BCR) wands, used with the GEM Premier 5000 and GEM Premier 7000 systems, to assign patient results to an incorrect patient ID. The recall status is ongoing. Affected quantity: 3.

Recall details

Product Type
Devices
Report Date
Aug 5, 2026
Initiation Date
May 8, 2026
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
3

Product description

Brand Name: GEM Premier 5000 Product Name: GEM Premier 5000 Model/Catalog Number: 00024019255 Portable blood analyzer system

Reason for recall

Potential for Bar Code Reader (BCR) wands, used with the GEM Premier 5000 and GEM Premier 7000 systems, to assign patient results to an incorrect patient ID.

Distribution pattern

US Nationwide distribution.

Code information

GEM Premier 5000: Part number: 00024019255 UDI: 08426950870227 Units distributed in Argentina: GEM Premier 5000 Serial Nos.: 25100466, 24097933 and 23064897. Affected Accessory: Bar Code Reader (BCR) Wand that received a configuration update on or after 8 April 2026. Part No. 00024015859; UDI No. 08426950869634

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-2826-2026
openFDA file date
Nov 10, 2026
Loaded by RegDataLab
Nov 11, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Source: openFDA record for Z-2826-2026. Report a data issue: hello@regdatalab.com.