Z-0002-2027 Class I Ongoing

Recalled by Instrumentation Laboratory — Bedford, MA

FDA device recall Z-0002-2027 was initiated by Instrumentation Laboratory on Sep 3, 2026 and is designated Class I. Reason for recall: Potential for invalid calibration curve or negative mean bias of up to 47% with patient samples. The recall status is ongoing. Affected quantity: 1,414 units.

Recall Details

Product Type
Devices
Report Date
Oct 21, 2026
Initiation Date
Sep 3, 2026
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
1,414 units

Product Description

HemoslL AcuStar HIT-lgG(PF-4). Part Number: 0009802028. in vitro diagnostic (OUS only)

Reason for Recall

Potential for invalid calibration curve or negative mean bias of up to 47% with patient samples.

Distribution Pattern

Worldwide - US Nationwide distribution in the states of AR, AZ, CA, CT, FL, GA, MA, MD, NJ, NY, PA, RI, TN, WA, WV and the countries of Austria, Belgium, Switzerland, Costa Rica, Czech Republic, Germany, Denmark, Spain, Finland, France, United Kingdom, Greece, India, Italy, Jordan, Japan, South Korea, Luxembourg, Latvia, Netherlands, Saudi Arabia, Sweden, Singapore.

Code Information

Part Number: 0009802028. UDI-DI: 08426950545774. Lot Number (Expiration Date): B40006 (01/16/2027), B38063 (11/01/2026).

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-0002-2027
openFDA file date
Oct 26, 2026
Loaded by RegDataLab
Oct 27, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.