Z-2835-2018 Class II Ongoing
Recall Details
- Product Type
- Devices
- Report Date
- August 22, 2018
- Initiation Date
- May 17, 2018
- Termination Date
- N/A
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 1
Product Description
The da Vinci Xi Surgical System (IS4000) Model Name: IS4000, Endoscope Controller. Model Number: 372601-16.
Reason for Recall
An Endoscope Controller has been improperly calibrated during manufacturing. This improper calibration may result in the Endoscope Controller showing surgical images that are dimmer, have color imbalance or have reduced fluorescence and grayscale when using Firefly. There is also potential for increased fogging of the endoscope.
Distribution Pattern
U.S.
Code Information
UDI Number: 00886874114193. Serial Number: SK1703.