Z-2838-2018 Class II Terminated

Recalled by Siemens Medical Solutions USA, Inc — Malvern, PA

FDA device recall Z-2838-2018 was initiated by Siemens Medical Solutions USA, Inc on Jun 29, 2018 and is designated Class II. Reason for recall: In rare cases of insufficient maintenance or high clinical workload, the first rope of the ceiling stand, which is designed to take the load, could break without triggering the safety lock and this… The recall status is terminated (terminated Oct 4, 2019). Affected quantity: 667.

Recall Details

Product Type
Devices
Report Date
Aug 29, 2018
Initiation Date
Jun 29, 2018
Termination Date
Oct 4, 2019
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
667

Product Description

3D V, model. no. 7042042 7042059 3070021 Product Usage: The ceiling stand is used to hold the X-ray tube and / or the detector and is moveable on a ceiling rail system in three axis. The tube and / or detector can be rotated above a horizontal and a vertical axle. The ceiling stand can be used within a stationary diagnostic X-ray system as well.

Reason for Recall

In rare cases of insufficient maintenance or high clinical workload, the first rope of the ceiling stand, which is designed to take the load, could break without triggering the safety lock and this could lead to overloading the safety rope which is not designed for continuous load and ongoing movement under load. This may lead to mechanical fatigue and cause the arm to drop down during patient positioning. Patients and users may be seriously injured.

Distribution Pattern

US Nationwide Distribution to accounts in IL, FL, and NY.

Code Information

7042042 : S/N 1002 to 1787 7042059: S/N 1002 to 1921

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-2838-2018
openFDA file date
Oct 23, 2026
Loaded by RegDataLab
Oct 26, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.