Z-2875-2018 Class II Terminated

Recalled by Zimmer Biomet, Inc. — Warsaw, IN

Recall Details

Product Type
Devices
Report Date
August 29, 2018
Initiation Date
June 18, 2018
Termination Date
March 26, 2020
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
247 units

Product Description

Connecting Bolt, Model 14-442093

Reason for Recall

Product was manufactured from the incorrect material, which shows a potential decrease in the fatigue strength of the connecting bolt that can lead to fracture.

Distribution Pattern

AR, AZ, CA, FL, GA, IN, KY, LA, MA, MI, MN, MO, MT, NJ, NY, OH, SC, TX, UT & WI

Code Information

...