Z-2876-2018 Class II Terminated

Recalled by Zimmer Biomet, Inc. — Warsaw, IN

Recall Details

Product Type
Devices
Report Date
August 29, 2018
Initiation Date
June 22, 2018
Termination Date
October 9, 2019
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
385

Product Description

Persona Partial Knee System Spacer Block Alignment Tower, Item No. 42-5399-006-35 Product Usage: Intended to aid in preparatory surgical procedures

Reason for Recall

The PPK System Spacer Block Alignment Tower mating feature that mates with the spacer block was undersized, which could potentially lead to mating issues between the devices. This could cause a delay in surgery of less than 30 minutes. There is an option in the surgical technique for performing limb alignment without the alignment tower.

Distribution Pattern

Worldwide Distribution - Nationwide distribution and the countries of Japan and Netherlands.

Code Information

63465803, 63762260, 63767434, 63799794