Z-2923-2018 Class II Terminated
FDA device recall Z-2923-2018 was initiated by Tecan US, Inc. on Jan 12, 2018 and is designated Class II. Reason for recall: Product is not meeting specification for traceability to NIST cortisol standard, generating 1.3-1.8 times higher cortisol values with laboratory samples in comparison to NIST cortisol standard. The recall status is terminated (terminated Jun 2, 2020). Affected quantity: 870 (USA) 180 (OUS).
Recall details
- Product Type
- Devices
- Report Date
- Sep 5, 2018
- Initiation Date
- Jan 12, 2018
- Termination Date
- Jun 2, 2020
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 870 (USA) 180 (OUS)
Product description
Reason for recall
Distribution pattern
Worldwide distribution; US distribution to states of: GA, IA, LA, MA, MN, OR, PA, SC, and UT; and countries of: Germany, Finland, Switzerland, France, Austria, Denmark, Poland, Norway, Belgium, Netherlands, Italy, Canada, South Africa, Russia, and Japan.
Code information
Source
- Dataset
- openFDA device recall enforcement reports
- FDA record
- This recall in openFDA (JSON)
- Enforcement Report
- Search FDA's Enforcement Report for Z-2923-2018
- openFDA file date
- Nov 19, 2026
- Loaded by RegDataLab
- Nov 20, 2026