Z-1967-2015 Class II Terminated

Recalled by Tecan US, Inc. — Morrisville, NC

FDA device recall Z-1967-2015 was initiated by Tecan US, Inc. on May 29, 2015 and is designated Class II. Reason for recall: A firmware bug causes incorrect values to be displayed for specific measured luminescence values. The recall status is terminated (terminated Aug 18, 2017). Affected quantity: 868 in total.

Recall details

Product Type
Devices
Report Date
Jul 8, 2015
Initiation Date
May 29, 2015
Termination Date
Aug 18, 2017
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
868 in total

Product description

Infinite F500, in vitro diagnostic. Product Usage: The Tecan Infinite F500 is a multifunctional microplate reader with injector option. The Infinite F500 provides high performance for the vast majority of todays microplate applications and research and is robotic compatible. The Infinite F500 is intended as a general purpose laboratory instrument for professional use, supporting common microplate conforming to the ANS/SBS standards.

Reason for recall

A firmware bug causes incorrect values to be displayed for specific measured luminescence values.

Distribution pattern

US Nationwide Distribution in the states of AL, AK, AZ, AR, CA, CO, CT, DE, FL, GA, IL, IN, IA, KS, KY, LA, ME, MD, MA, MI, MN, MO, NE NH, NJ, NM, NY, NC, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, WA, WI, including DC

Code information

Material Number 30019337. Serial numbers 1212004416 1009002605 612000003 811002510 712004158 906000965 902005109 912003472 1103002583 1411007090 704000003 702000001 708000500 1003004146 1109000103 1412005124 805001219 911005026 708005532 705007195 810000248 1010001663 712004152 708005533 1412005709 1103003780 706004455 608000002 908000185 1009002604 1308001035 804005914 1311007600 804000984 705009034 1309003195 801008055 706007961 901006341

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-1967-2015
openFDA file date
Nov 19, 2026
Loaded by RegDataLab
Nov 20, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Need the full FDA record for Tecan US, Inc.? Ask Claude or ChatGPT with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. The API key is free: no card.

Get a free API key

How the connector works

Source: openFDA record for Z-1967-2015. Report a data issue: hello@regdatalab.com.