Z-2928-2018 Class II Terminated

Recalled by Siemens Healthcare Diagnostics, Inc. — Newark, DE

FDA device recall Z-2928-2018 was initiated by Siemens Healthcare Diagnostics, Inc. on Jul 16, 2018 and is designated Class II. Reason for recall: Therapeutic levels (1mg/mL and 2mg/mL) of Hydroxocobalamin may interfere with tHb and some of the CO-Ox fractions that are reported on the RAPIDPoint and RAPIDLab Blood Gas systems indicated. This may… The recall status is terminated (terminated Jun 23, 2020). Affected quantity: 10199.

Recall Details

Product Type
Devices
Report Date
Sep 5, 2018
Initiation Date
Jul 16, 2018
Termination Date
Jun 23, 2020
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
10199

Product Description

RAPIDPoint 500 Blood Gas Analyzer; 10492730, 1069855, 10696857, 10697306 Device is point-of-care system for comprehensive lab-quality testing to patient's bedside. Test menu includes blood gas, electrolytes, glucose, lactate, neonatal total bilirubin and full CO-oximetry including total hemoglobin.

Reason for Recall

Therapeutic levels (1mg/mL and 2mg/mL) of Hydroxocobalamin may interfere with tHb and some of the CO-Ox fractions that are reported on the RAPIDPoint and RAPIDLab Blood Gas systems indicated. This may cause lower than expected values for carboxyhemoglobin (fCOHb) and methemoglobin (fMetHb). A negative interference with fCOHb has the potential to alter the medical assessment of the patient and may withhold necessary follow-up treatment and/or initiate cessation of medication in response to elevated fMetHb levels.

Distribution Pattern

Nationwide distribution. Worldwide foreign distribution.

Code Information

All units affected

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-2928-2018
openFDA file date
Oct 23, 2026
Loaded by RegDataLab
Oct 25, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.