Z-2929-2018 Class II Terminated

Recalled by Siemens Healthcare Diagnostics, Inc. — Newark, DE

Recall Details

Product Type
Devices
Report Date
September 5, 2018
Initiation Date
July 16, 2018
Termination Date
June 23, 2020
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
138

Product Description

RAPIDLab 1245 Blood Gas Analyzer; 10321844, 10337179, 10491393 Device is point-of-care system for comprehensive lab-quality testing to patient's bedside. Test menu includes blood gas, electrolytes, glucose, lactate, neonatal total bilirubin and full CO-oximetry including total hemoglobin.

Reason for Recall

Therapeutic levels (1mg/mL and 2mg/mL) of Hydroxocobalamin may interfere with tHb and some of the CO-Ox fractions that are reported on the RAPIDPoint and RAPIDLab Blood Gas systems indicated. This may cause lower than expected values for carboxyhemoglobin (fCOHb) and methemoglobin (fMetHb). A negative interference with fCOHb has the potential to alter the medical assessment of the patient and may withhold necessary follow-up treatment and/or initiate cessation of medication in response to elevated fMetHb levels.

Distribution Pattern

Nationwide distribution. Worldwide foreign distribution.

Code Information

All units affected