Z-2941-2020 Class I Terminated

Recalled by MicroPort Orthopedics Inc. — Arlington, TN

FDA device recall Z-2941-2020 was initiated by MicroPort Orthopedics Inc. on Jul 31, 2020 and is designated Class I. Reason for recall: MicroPort Orthopedics Inc. states that there have been reports of fractures of the long and extra-long Titanium modular femoral neck component after implantation. Worldwide there were 680 Long and… The recall status is terminated (terminated Apr 16, 2024). Affected quantity: 123,284 units.

Recall details

Product Type
Devices
Report Date
Sep 30, 2020
Initiation Date
Jul 31, 2020
Termination Date
Apr 16, 2024
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
123,284 units

Product description

PROFEMUR Neck Long Titanium, Catalog Numbers: PHA01204, PHA01214, PHA01224, PHA01234, PHA01244, PHA01254, PHA01264. Hip prosthesis component.

Reason for recall

MicroPort Orthopedics Inc. states that there have been reports of fractures of the long and extra-long Titanium modular femoral neck component after implantation. Worldwide there were 680 Long and X-Long PROFEMUR Titanium modular neck fractures among 123,840 global sales representing is a cumulative rate of modular neck device fracture of 0.55% from 2002 to 2019. Within the United States there were 349 Long and X-Long PROFEMUR¿ Titanium modular neck fractures among 15,786 United States sales representing a cumulative rate of modular neck device fracture with a fracture rate in the United States of 2.21% from 2002 to 2019. There is an average time of 5.4 years to device fracture after implantation.

Distribution pattern

worldwide, except China and Japan

Code information

All lots/serial numbers

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-2941-2020
openFDA file date
Nov 10, 2026
Loaded by RegDataLab
Nov 11, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Source: openFDA record for Z-2941-2020. Report a data issue: hello@regdatalab.com.