Z-2941-2020 Class I Terminated

Recalled by MicroPort Orthopedics Inc. — Arlington, TN

Recall Details

Product Type
Devices
Report Date
September 30, 2020
Initiation Date
July 31, 2020
Termination Date
April 16, 2024
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
123,284 units

Product Description

PROFEMUR Neck Long Titanium, Catalog Numbers: PHA01204, PHA01214, PHA01224, PHA01234, PHA01244, PHA01254, PHA01264. Hip prosthesis component.

Reason for Recall

MicroPort Orthopedics Inc. states that there have been reports of fractures of the long and extra-long Titanium modular femoral neck component after implantation. Worldwide there were 680 Long and X-Long PROFEMUR Titanium modular neck fractures among 123,840 global sales representing is a cumulative rate of modular neck device fracture of 0.55% from 2002 to 2019. Within the United States there were 349 Long and X-Long PROFEMUR¿ Titanium modular neck fractures among 15,786 United States sales representing a cumulative rate of modular neck device fracture with a fracture rate in the United States of 2.21% from 2002 to 2019. There is an average time of 5.4 years to device fracture after implantation.

Distribution Pattern

worldwide, except China and Japan

Code Information

All lots/serial numbers