Z-2942-2020 Class I Terminated
FDA device recall Z-2942-2020 was initiated by MicroPort Orthopedics Inc. on Jul 31, 2020 and is designated Class I. Reason for recall: MicroPort Orthopedics Inc. states that there have been reports of fractures of the long and extra-long Titanium modular femoral neck component after implantation. Worldwide there were 680 Long and… The recall status is terminated (terminated Apr 16, 2024). Affected quantity: 556 units.
Recall details
- Product Type
- Devices
- Report Date
- Sep 30, 2020
- Initiation Date
- Jul 31, 2020
- Termination Date
- Apr 16, 2024
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 556 units
Product description
PROFEMUR Neck Extra Long Titanium, Catalog Numbers: PHA01206, PHA01236, PHA01256. Hip prosthesis component
Reason for recall
MicroPort Orthopedics Inc. states that there have been reports of fractures of the long and extra-long Titanium modular femoral neck component after implantation. Worldwide there were 680 Long and X-Long PROFEMUR Titanium modular neck fractures among 123,840 global sales representing is a cumulative rate of modular neck device fracture of 0.55% from 2002 to 2019. Within the United States there were 349 Long and X-Long PROFEMUR Titanium modular neck fractures among 15,786 United States sales representing a cumulative rate of modular neck device fracture with a fracture rate in the United States of 2.21% from 2002 to 2019. There is an average time of 5.4 years to device fracture after implantation.
Distribution pattern
worldwide, except China and Japan
Code information
All lots/serial numbers
Source
- Dataset
- openFDA device recall enforcement reports
- FDA record
- This recall in openFDA (JSON)
- Enforcement Report
- Search FDA's Enforcement Report for Z-2942-2020
- openFDA file date
- Nov 10, 2026
- Loaded by RegDataLab
- Nov 11, 2026