Z-2951-2024 Class II Ongoing

Recalled by Mindray DS USA, Inc. dba Mindray North America — Mahwah, NJ

Recall Details

Product Type
Devices
Report Date
September 11, 2024
Initiation Date
July 22, 2024
Termination Date
N/A
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
24 units

Product Description

Isoflurane E_Vap.(Safety Filling adapter): used with the A9 Anesthesia System-Used to provide the vapor of an anesthetic agent with controllable concentration to an A9 anesthesia system. Part Number: Part Number 115-066758-00

Reason for Recall

V90 electronic vaporizer anesthetic agent output may be higher than the set amount due to a component within the vaporizer causing an incomplete seal in some cases. Should this issue occur, the A9 Anesthesia System will generate alarms associated with abnormal vaporizer output. If the anesthetic concentration is too high, the patient can inhale too much anesthetic agent; the patient may suffer from symptoms such as vertigo and drowsiness after the operation. In severe cases, the patient may suffer from neurological sequelae such as hypomnesia and inattention

Distribution Pattern

US Distribution to states of: California, Georgia, Kentucky, Mississippi, New Jersey, Puerto Rico, Texas, and West Virginia; and International (Foreign) to: Canada

Code Information

UDI-DI: 06936415975753 Serial Number: AH6B24000165 AH6B24000170 AH6B24000167 AH6B24000163 AH6B24000166 AH6B24000173 AH6B24000174 AH6B24000183