Z-2949-2024 Class II Ongoing
FDA device recall Z-2949-2024 was initiated by Mindray DS USA, Inc. Dba Mindray North America on Jul 22, 2024 and is designated Class II. Reason for recall: V90 electronic vaporizer anesthetic agent output may be higher than the set amount due to a component within the vaporizer causing an incomplete seal in some cases. Should this issue occur, the A9… The recall status is ongoing. Affected quantity: 2 units.
Recall details
- Product Type
- Devices
- Report Date
- Sep 11, 2024
- Initiation Date
- Jul 22, 2024
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 2 units
Product description
Sevoflurane E_Vap.(Quik-Fil):used with the A9 Anesthesia System-Used to provide the vapor of an anesthetic agent with controllable concentration to an A9 anesthesia system. Part Number 115-066755-00
Reason for recall
V90 electronic vaporizer anesthetic agent output may be higher than the set amount due to a component within the vaporizer causing an incomplete seal in some cases. Should this issue occur, the A9 Anesthesia System will generate alarms associated with abnormal vaporizer output. If the anesthetic concentration is too high, the patient can inhale too much anesthetic agent; the patient may suffer from symptoms such as vertigo and drowsiness after the operation. In severe cases, the patient may suffer from neurological sequelae such as hypomnesia and inattention
Distribution pattern
US Distribution to states of: California, Georgia, Kentucky, Mississippi, New Jersey, Puerto Rico, Texas, and West Virginia; and International (Foreign) to: Canada
Code information
UDI-DI: 06936415975784 Serial Number: AH6D28000517, AH6D28000518