Z-2996-2018 Class I Terminated
Recall Details
- Product Type
- Devices
- Report Date
- September 26, 2018
- Initiation Date
- August 2, 2018
- Termination Date
- November 22, 2021
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 11,712 units
Product Description
OptiView DAB IHC Detection Kit, Catalog Number 06396500001, model 760-700 Immunohistochemistry (IHC) for in vitro diagnostic use.
Reason for Recall
Leaking and sticking reagent dispensers for horseradish peroxidase (HRP) dispensers. A false negative IHC result may occur.
Distribution Pattern
worldwide
Code Information
Lot Numbers: Y25760, E00119, Y24225, UDI: 4015630984039