Z-3008-2018 Class I Terminated

Recalled by Ventana Medical Systems Inc — Oro Valley, AZ

FDA device recall Z-3008-2018 was initiated by Ventana Medical Systems Inc on Aug 2, 2018 and is designated Class I. Reason for recall: Leaking and sticking reagent dispensers for horseradish peroxidase (HRP) dispensers The recall status is terminated (terminated Nov 22, 2021). Affected quantity: 559 units.

Recall Details

Product Type
Devices
Report Date
Sep 26, 2018
Initiation Date
Aug 2, 2018
Termination Date
Nov 22, 2021
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
559 units

Product Description

VENTANA ISH iView Blue Detection Kit, Catalog Number 05278511001, model 800-092 Immunohistochemistry (IHC) for in vitro diagnostic use.

Reason for Recall

Leaking and sticking reagent dispensers for horseradish peroxidase (HRP) dispensers

Distribution Pattern

worldwide

Code Information

Lot Numbers: Y15105, Y22455, UDI: 4015630971930

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-3008-2018
openFDA file date
Oct 6, 2026
Loaded by RegDataLab
Oct 12, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it. FDA may have changed the recall since.