Z-3113-2024 Class II Ongoing

Recalled by MEDLINE INDUSTRIES, LP - Northfield — Northfield, IL

FDA device recall Z-3113-2024 was initiated by MEDLINE INDUSTRIES, LP - Northfield on Jun 27, 2024 and is designated Class II. Reason for recall: Due to reported adverse events, Medline Industries, LP. is issuing a recall on specific SKUs and lots of 1500 cc semi-rigid liners manufactured from 09/2023-11/2023 due to the potential loss of… The recall status is ongoing. Affected quantity: 350 units.

Recall Details

Product Type
Devices
Report Date
Sep 18, 2024
Initiation Date
Jun 27, 2024
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
350 units

Product Description

Medline convenience kits containing semi-rigid suction liners labeled as follows: a) MINOR BREAST, Pack Number DYNJ908104B; b) PLASTIC MAJOR, Pack Number DYNJ908106B; c) ABDOMINOPLASTY W/LIPO SUCTION, Pack Number DYNJ909946; d) PLASTIC, Pack Number DYNJ909951

Reason for Recall

Due to reported adverse events, Medline Industries, LP. is issuing a recall on specific SKUs and lots of 1500 cc semi-rigid liners manufactured from 09/2023-11/2023 due to the potential loss of suction or low suction during intermittent suction use.

Distribution Pattern

Worldwide distribution - US Nationwide and the countries of Canada, Panama.

Code Information

a) DYNJ908104B: UDI/DI case 40195327507115, UDI/DI each 10195327507114, Lot Numbers: 23LBT869, 24BBQ730; b) DYNJ908106B: UDI/DI case 40195327507139, UDI/DI each 10195327507138, Lot Numbers: 24ABB087, 24CBM739, 24EBH234; c) DYNJ909946: UDI/DI case 40195327512652, UDI/DI each 10195327512651, Lot Numbers: 23LBD934, 24DBB806; d) DYNJ909951: UDI/DI case 40195327512904, UDI/DI each 10195327512903, Lot Numbers: 23LBD932, 24ABP740

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-3113-2024
openFDA file date
Oct 23, 2026
Loaded by RegDataLab
Oct 25, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.