Z-3119-2024 Class II Ongoing
Recall Details
- Product Type
- Devices
- Report Date
- September 18, 2024
- Initiation Date
- June 27, 2024
- Termination Date
- N/A
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 35 units
Product Description
Medline convenience kits containing semi-rigid suction liners labeled as follows: MERCY TIFFIN C-SECTION, Pack Number DYNJ909732
Reason for Recall
Due to reported adverse events, Medline Industries, LP. is issuing a recall on specific SKUs and lots of 1500 cc semi-rigid liners manufactured from 09/2023-11/2023 due to the potential loss of suction or low suction during intermittent suction use.
Distribution Pattern
Worldwide distribution - US Nationwide and the countries of Canada, Panama.
Code Information
DYNJ909732: UDI/DI case 4019532746469240195327464746, UDI/DI each 10195327464745, Lot Numbers: 23LBT911, 24ABX052, 24CBE496