Z-3179-2024 Class II Ongoing

Recalled by B Braun Medical Inc — Bethlehem, PA

Recall Details

Product Type
Devices
Report Date
September 25, 2024
Initiation Date
August 15, 2024
Termination Date
N/A
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
11262112 units

Product Description

STREAMLINE BLOODLINE SET FOR DIALOG DR-Tubing to be used to transfer blood between a patient and a haemodialyser during haemodialysis. Model/Catalog Number: SL-2010M2096A

Reason for Recall

Incomplete insertion of the patient connector of the Streamline Airless System Hemodialysis Bloodline to the Locksite may result in the potential for occlusion during prograde rinseback procedures. If the patient connector is not fully inserted into the Locksite, the internal septum may not open fully enough to prevent restrictions to flow. may cause delays in therapy while clinicians are forced to switch to retrograde flow method

Distribution Pattern

US Nationwide distribution.

Code Information

UDI-DI 04046955674992; All Unexpired Lot Numbers