Z-3179-2024 Class II Ongoing

Recalled by B Braun Medical Inc — Bethlehem, PA

FDA device recall Z-3179-2024 was initiated by B Braun Medical Inc on Aug 15, 2024 and is designated Class II. Reason for recall: Incomplete insertion of the patient connector of the Streamline Airless System Hemodialysis Bloodline to the Locksite may result in the potential for occlusion during prograde rinseback procedures… The recall status is ongoing. Affected quantity: 11262112 units.

Recall Details

Product Type
Devices
Report Date
Sep 25, 2024
Initiation Date
Aug 15, 2024
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
11262112 units

Product Description

STREAMLINE BLOODLINE SET FOR DIALOG DR-Tubing to be used to transfer blood between a patient and a haemodialyser during haemodialysis. Model/Catalog Number: SL-2010M2096A

Reason for Recall

Incomplete insertion of the patient connector of the Streamline Airless System Hemodialysis Bloodline to the Locksite may result in the potential for occlusion during prograde rinseback procedures. If the patient connector is not fully inserted into the Locksite, the internal septum may not open fully enough to prevent restrictions to flow. may cause delays in therapy while clinicians are forced to switch to retrograde flow method

Distribution Pattern

US Nationwide distribution.

Code Information

UDI-DI 04046955674992; All Unexpired Lot Numbers

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-3179-2024
openFDA file date
Oct 23, 2026
Loaded by RegDataLab
Oct 26, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.