Z-3201-2024 Class II Ongoing

Recalled by Philips North America Llc — Cambridge, MA

FDA device recall Z-3201-2024 was initiated by Philips North America Llc on Sep 3, 2024 and is designated Class II. Reason for recall: Integrated Radio Frequency (IRF) Carrier assembly screws may come loose and protrude beyond the surface could cause the screws to interfere with other moving parts, potentially blocking the horizontal… The recall status is ongoing. Affected quantity: 68 units.

Recall Details

Product Type
Devices
Report Date
Oct 2, 2024
Initiation Date
Sep 3, 2024
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
68 units

Product Description

MR 5300 - For use as a diagnostic device to obtain cross-sectional images, spectroscopic images and/or spectra of the internal structure of the head, body or extremities, in any orientation. (REF) Numbers: 782110

Reason for Recall

Integrated Radio Frequency (IRF) Carrier assembly screws may come loose and protrude beyond the surface could cause the screws to interfere with other moving parts, potentially blocking the horizontal tabletop. May lead to delay in diagnosis and/or may include anxiety of a patient is left on the tabletop inside the bore for a longer duration than anticipated. movement.

Distribution Pattern

Worldwide distribution - US Nationwide and the countries of Argentina, Australia, Belgium, Brazil, Canada, Chile, China, Colombia, Croatia, Czech Republic, Denmark, Ecuador, Egypt, Finland, France, Germany, Iceland, India, Indonesia, Iran, Iraq, Israel, Italy, Japan, Jordan, Kuwait, Latvia, Maldives, Mauritius, Mexico, Morocco, Netherlands, Pakistan, Peru, Poland, Portugal, Qatar, Romania, Russian Fed., Saudi Arabia, Singapore, Slovenia, South Africa, South Korea, Spain, Switzerland, Thailand, T¿rkiye, United Kingdom, Utd.Arab.Emir., Uzbekistan, Vietnam.

Code Information

UDI-DI: (01)00884838099364(21) Serial Numbers: 63027 63035 63039 63066 63068 63070 63076 63078 63083 63084 63086 63092 63094 63096 63097 63103 63104 63106 63109 63112 63113 63115 63122 63123 63125 63128 63130 63132 63135 63137 63140 63141 63142 63143 63144 63146 63148 63149 63150 63151 63152 63154 63157 63160 63161 63168 63169 63172 63174 63177 63178 63183 63186 63190 63191 63193 63197 63201 63203 63211 63229 63238 63248 63263 63272 63288 63367 63370

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-3201-2024
openFDA file date
Oct 23, 2026
Loaded by RegDataLab
Oct 25, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.