Z-3257-2026 Class II Ongoing
FDA device recall Z-3257-2026 was initiated by Baxter Healthcare Corporation on Aug 3, 2026 and is designated Class II. Reason for recall: Contamination within the release test equipment flow path could have affected the accuracy of the nebulizer flow rate. Affected devices returned to firm and testing confirmed devices meet… The recall status is ongoing. Affected quantity: 2.
Recall Details
- Product Type
- Devices
- Report Date
- Sep 30, 2026
- Initiation Date
- Aug 3, 2026
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 2
Product Description
Brand Name: VOLARA System Product Name: VOLARA, HC, NA A Model/Catalog Number: M08594A Software Version: N/A Product Description: VOLARA, HOMECARE, NORTH AMERICA Component: No
Reason for Recall
Contamination within the release test equipment flow path could have affected the accuracy of the nebulizer flow rate. Affected devices returned to firm and testing confirmed devices meet specifications.
Distribution Pattern
US Nationwide distribution in the states of Colorado and Florida.
Code Information
Lot Code: UDI: 00887761981690 Serial Numbers: W078SY3553 and Z309SY7853