Z-3281-2018 Class II Terminated

Recalled by Stryker Neurovascular — Fremont, CA

Recall Details

Product Type
Devices
Report Date
October 10, 2018
Initiation Date
May 2, 2018
Termination Date
N/A
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
15 units

Product Description

AXS Infinity LS Long Sheath:((6F/.088in)/ x 70cm GEN-10800-70, (6F/.088in)/ x 80cm GEN-10800-80, (6F/.088in)/ x 90cm GEN-10800-90) Made in Ireland, Sterile EO. The AXS Infinity LS Long Sheath system contents are: One (1) AXS Infinity LS Long Sheath, One (1) Hemostasis valve with extension tubing, One Long Dilator.

Reason for Recall

The AXS Infinity LS Long Sheath Dilator outer diameter (OD) specification was exceeded.

Distribution Pattern

Only U.S. Consignees: CO,NJ,NY, OH,PA, and WV.

Code Information

Lot No. 20156091 ;UPN: GEN-10800-80-S2 UDI (01)07613327298253 (17)201130 (10) 20156091.