FDA Inspection 1056379 — New Era Orthopaedics, LLc

New Era Orthopaedics, LLc — FEI 3010760160

The FDA completed an inspection of New Era Orthopaedics, LLc on Jun 11, 2018 in Lake Mary, FL. The inspection was classified VAI (Voluntary Action Indicated — objectionable conditions or practices were found, but the agency is not prepared to take or recommend any administrative or regulatory action). 11 citations were issued during this inspection. Product area: Devices.

Inspection Details

Classification
Voluntary Action Indicated (VAI)
End Date
Jun 11, 2018
Fiscal Year
2018
Product Type
Devices
Project Area
Postmarket Assurance: Devices
Location
Lake Mary, FL (United States)

Citations

IDCFRDescription
1471421 CFR 820.30(c)Design input - Lack of or inadequate procedures
1471621 CFR 820.30(f)Design verification - output does not meet input requirement
1472221 CFR 820.40Procedures not adequately established or maintained
310121 CFR 820.30(g)Design validation- Lack of or inadequate procedures
310221 CFR 820.30(h)Design transfer - Lack of or inadequate procedures
310821 CFR 820.70(e)Contamination control, Lack of or inadequate procedures
313021 CFR 820.100(a)Lack of or inadequate procedures
333121 CFR 820.181DMR - not or inadequately maintained
367821 CFR 820.30(g)Design Validation - Risk analysis not performed/inadequate
419321 CFR 806.10(b)Time to report - 10 days
47921 CFR 820.50Purchasing controls, Lack of or inadequate procedures