FDA Inspection 1074846 — Spring Health Products Inc

Spring Health Products Inc — FEI 2523990

The FDA completed an inspection of Spring Health Products Inc on Dec 18, 2018 in Norristown, PA. The inspection was classified VAI (Voluntary Action Indicated — objectionable conditions or practices were found, but the agency is not prepared to take or recommend any administrative or regulatory action). 13 citations were issued during this inspection. Product area: Devices.

Inspection Details

Classification
Voluntary Action Indicated (VAI)
End Date
Dec 18, 2018
Fiscal Year
2019
Product Type
Devices
Project Area
Compliance: Devices
Location
Norristown, PA (United States)

Citations

IDCFRDescription
1471221 CFR 820.184DHR - not or inadequately maintained
310321 CFR 820.30(i)Design changes - Lack of or Inadequate Procedures
310421 CFR 820.30(j)Design history file
312821 CFR 820.90(a)Nonconforming product control
313021 CFR 820.100(a)Lack of or inadequate procedures
313221 CFR 820.120Lack of or inadequate procedures for labeling
316821 CFR 820.198(a)Complaints
317221 CFR 820.198(c)Investigation of device failures
320621 CFR 820.50(b)Approval, inadequate purchasing data
323321 CFR 820.72(a)Calibration, Inspection, etc. Procedures Lack of or Inadequ
333121 CFR 820.181DMR - not or inadequately maintained
367821 CFR 820.30(g)Design Validation - Risk analysis not performed/inadequate
405921 CFR 820.22Quality Audits - defined intervals