FDA Inspection 1166872 — Zuga Medical Inc.
The FDA completed an inspection of Zuga Medical Inc. on April 12, 2022 in Chagrin Falls, OH. The inspection was classified VAI (Voluntary Action Indicated — objectionable conditions were found but do not meet the threshold for regulatory action). 8 citations were issued during this inspection. Product area: Devices.
Inspection Details
- Classification
- Voluntary Action Indicated (VAI) (VAI)
- End Date
- April 12, 2022
- Fiscal Year
- 2022
- Product Type
- Devices
- Project Area
- Postmarket Assurance: Devices
- Location
- Chagrin Falls, OH (United States)
Citations
| ID | CFR | Description |
|---|---|---|
| 2327 | 21 CFR 820.22 | Quality audits - Lack of or inadequate procedures |
| 3102 | 21 CFR 820.30(h) | Design transfer - Lack of or inadequate procedures |
| 3103 | 21 CFR 820.30(i) | Design changes - Lack of or Inadequate Procedures |
| 3130 | 21 CFR 820.100(a) | Lack of or inadequate procedures |
| 3168 | 21 CFR 820.198(a) | Complaints |
| 3669 | 21 CFR 820.20(c) | Management review - defined interval, sufficient frequency |
| 3678 | 21 CFR 820.30(g) | Design Validation - Risk analysis not performed/inadequate |
| 479 | 21 CFR 820.50 | Purchasing controls, Lack of or inadequate procedures |