FDA Inspection 880115 — Zuga Medical Inc.
The FDA completed an inspection of Zuga Medical Inc. on May 16, 2014 in Chagrin Falls, OH. The inspection was classified VAI (Voluntary Action Indicated — objectionable conditions were found but do not meet the threshold for regulatory action). 7 citations were issued during this inspection. Product area: Devices.
Inspection Details
- Classification
- Voluntary Action Indicated (VAI) (VAI)
- End Date
- May 16, 2014
- Fiscal Year
- 2014
- Product Type
- Devices
- Project Area
- Postmarket Assurance: Devices
- Location
- Chagrin Falls, OH (United States)
Citations
| ID | CFR | Description |
|---|---|---|
| 14716 | 21 CFR 820.30(f) | Design verification - output does not meet input requirement |
| 2557 | 21 CFR 820.30(c) | Design input - documentation |
| 3101 | 21 CFR 820.30(g) | Design validation- Lack of or inadequate procedures |
| 3282 | 21 CFR 820.90(a) | Nonconforming product, Lack of or inadequate procedures |
| 479 | 21 CFR 820.50 | Purchasing controls, Lack of or inadequate procedures |
| 486 | 21 CFR 820.50(a) | Evaluation of suppliers, contractors, etc., requirements |
| 630 | 21 CFR 803.17 | Lack of Written MDR Procedures |