FDA Inspection 880115 — Zuga Medical Inc.

Zuga Medical Inc. — FEI 3010251542

The FDA completed an inspection of Zuga Medical Inc. on May 16, 2014 in Chagrin Falls, OH. The inspection was classified VAI (Voluntary Action Indicated — objectionable conditions were found but do not meet the threshold for regulatory action). 7 citations were issued during this inspection. Product area: Devices.

Inspection Details

Classification
Voluntary Action Indicated (VAI) (VAI)
End Date
May 16, 2014
Fiscal Year
2014
Product Type
Devices
Project Area
Postmarket Assurance: Devices
Location
Chagrin Falls, OH (United States)

Citations

IDCFRDescription
1471621 CFR 820.30(f)Design verification - output does not meet input requirement
255721 CFR 820.30(c)Design input - documentation
310121 CFR 820.30(g)Design validation- Lack of or inadequate procedures
328221 CFR 820.90(a)Nonconforming product, Lack of or inadequate procedures
47921 CFR 820.50Purchasing controls, Lack of or inadequate procedures
48621 CFR 820.50(a)Evaluation of suppliers, contractors, etc., requirements
63021 CFR 803.17Lack of Written MDR Procedures