FDA Inspection 1182299 — Primrose Medical Inc.

Primrose Medical Inc. — FEI 1220905

The FDA completed an inspection of Primrose Medical Inc. on Oct 26, 2022 in East Walpole, MA. The inspection was classified VAI (Voluntary Action Indicated — objectionable conditions or practices were found, but the agency is not prepared to take or recommend any administrative or regulatory action). 11 citations were issued during this inspection. Product area: Devices.

Inspection Details

Classification
Voluntary Action Indicated (VAI)
End Date
Oct 26, 2022
Fiscal Year
2023
Product Type
Devices
Project Area
Postmarket Assurance: Devices
Location
East Walpole, MA (United States)

Citations

IDCFRDescription
1471321 CFR 820.198(a)Lack of or inadequate complaint procedures
2142421 CFR 820.120(b)Labeling inspection - UDI
232821 CFR 820.22Quality audits - auditor independence
237121 CFR 820.30(a)Design control - no procedures
310421 CFR 820.30(j)Design history file
313021 CFR 820.100(a)Lack of or inadequate procedures
367821 CFR 820.30(g)Design Validation - Risk analysis not performed/inadequate
369621 CFR 820.100(b)Documentation
383721 CFR 820.25(b)Training records
47921 CFR 820.50Purchasing controls, Lack of or inadequate procedures
48621 CFR 820.50(a)Evaluation of suppliers, contractors, etc., requirements