FDA Inspection 1233779 — Revian Inc.

Revian Inc. — FEI 3015249767

The FDA completed an inspection of Revian Inc. on Mar 22, 2024 in Morrisville, NC. The inspection was classified NAI (No Action Indicated — no objectionable conditions or practices were found during the inspection). 8 citations were issued during this inspection. Product area: Devices.

Inspection Details

Classification
No Action Indicated (NAI)
End Date
Mar 22, 2024
Fiscal Year
2024
Product Type
Devices
Project Area
Postmarket Assurance: Devices
Location
Morrisville, NC (United States)

Citations

IDCFRDescription
1471621 CFR 820.30(f)Design verification - output does not meet input requirement
2280021 CFR 830.300(a)Devices subject to device identification GUDID data submission requirements.
313021 CFR 820.100(a)Lack of or inadequate procedures
316821 CFR 820.198(a)Complaints
323521 CFR 820.72(a)Equipment control activity documentation
328221 CFR 820.90(a)Nonconforming product, Lack of or inadequate procedures
367821 CFR 820.30(g)Design Validation - Risk analysis not performed/inadequate
48621 CFR 820.50(a)Evaluation of suppliers, contractors, etc., requirements