FDA Inspection 1234615 — Brace Align

Brace Align — FEI 3013002868

The FDA completed an inspection of Brace Align on Mar 29, 2024 in Soddy Daisy, TN. The inspection was classified VAI (Voluntary Action Indicated — objectionable conditions or practices were found, but the agency is not prepared to take or recommend any administrative or regulatory action). 12 citations were issued during this inspection. Product area: Devices.

Inspection Details

Classification
Voluntary Action Indicated (VAI)
End Date
Mar 29, 2024
Fiscal Year
2024
Product Type
Devices
Project Area
Postmarket Assurance: Devices
Location
Soddy Daisy, TN (United States)

Citations

IDCFRDescription
1471321 CFR 820.198(a)Lack of or inadequate complaint procedures
2279521 CFR 801.20(a)Label to bear a unique device identifier
237121 CFR 820.30(a)Design control - no procedures
312021 CFR 820.80(a)Lack of or inadequate procedures - Acceptance activities
313021 CFR 820.100(a)Lack of or inadequate procedures
316021 CFR 820.184Lack of or inadequate DHR procedures
328221 CFR 820.90(a)Nonconforming product, Lack of or inadequate procedures
333121 CFR 820.181DMR - not or inadequately maintained
366521 CFR 820.20(b)(3)Management representative
44721 CFR 820.40Lack of procedures, or not maintained
47921 CFR 820.50Purchasing controls, Lack of or inadequate procedures
63021 CFR 803.17Lack of Written MDR Procedures