FDA Inspection 1314396 — Pensar Medical

Pensar Medical — FEI 3007793449

The FDA completed an inspection of Pensar Medical on Apr 28, 2026 in Houston, TX. The inspection was classified OAI (Official Action Indicated — regulatory and/or administrative actions are recommended). 7 citations were issued during this inspection. Product area: Devices.

Inspection Details

Classification
Official Action Indicated (OAI)
End Date
Apr 28, 2026
Fiscal Year
2026
Product Type
Devices
Project Area
Compliance: Devices
Location
Houston, TX (United States)

Citations

IDCFRDescription
23019ISO 13485:2016 Clause 4.1.2Risk-based approach not applied to QMS processes
23109ISO 13485:2016 Clause 7.3.4Records of design and development outputs not maintained
23118ISO 13485:2016 Clause 7.3.6Records of design and development verification not maintained
23133ISO 13485:2016 Clause 7.3.9Requirements for design and development changes not met before implementation
23215ISO 13485:2016 Clause 8.2.2Complaint handling procedures lack required elements
23258ISO 13485:2016 Clause 8.5.2Corrective action procedures not documented
63121 CFR 803.17(a)(1)Lack of System for Event Evaluations