FDA Inspection 1320072 — Guardian Medical USA Inc.

Guardian Medical USA Inc. — FEI 3016261131

The FDA completed an inspection of Guardian Medical USA Inc. on Jun 2, 2026 in Swanton, OH. The inspection was classified VAI (Voluntary Action Indicated — objectionable conditions or practices were found, but the agency is not prepared to take or recommend any administrative or regulatory action). 6 citations were issued during this inspection. Product area: Devices.

Inspection Details

Classification
Voluntary Action Indicated (VAI)
End Date
Jun 2, 2026
Fiscal Year
2026
Product Type
Devices
Project Area
Compliance: Devices
Location
Swanton, OH (United States)

Citations

IDCFRDescription
23019ISO 13485:2016 Clause 4.1.2Risk-based approach not applied to QMS processes
23127ISO 13485:2016 Clause 7.3.8Design and development transfer procedures not documented
23139ISO 13485:2016 Clause 7.4.1Criteria for the evaluation and selection of suppliers not established
23168ISO 13485:2016 Clause 7.5.6Planned results not consistently met during validation of processes
23237ISO 13485:2016 Clause 8.3.1Procedure for nonconforming product not documented
23258ISO 13485:2016 Clause 8.5.2Corrective action procedures not documented