FDA Inspection 556414 — ITC Nexus Dx

ITC Nexus Dx — FEI 3005776351

The FDA completed an inspection of ITC Nexus Dx on Jan 7, 2009 in Piscataway, NJ. The inspection was classified OAI (Official Action Indicated — regulatory and/or administrative actions are recommended). 35 citations were issued during this inspection. Product area: Devices.

Inspection Details

Classification
Official Action Indicated (OAI)
End Date
Jan 7, 2009
Fiscal Year
2009
Product Type
Devices
Project Area
Postmarket Assurance: Devices
Location
Piscataway, NJ (United States)

Citations

IDCFRDescription
232721 CFR 820.22Quality audits - Lack of or inadequate procedures
235021 CFR 820.25(b)Training - Lack of or inadequate procedures
237121 CFR 820.30(a)Design control - no procedures
312121 CFR 820.80(b)Lack of or inadequate receiving acceptance procedures
312521 CFR 820.80(d)Lack of or inadequate final acceptance procedures
313021 CFR 820.100(a)Lack of or inadequate procedures
316021 CFR 820.184Lack of or inadequate DHR procedures
316421 CFR 820.184(d)Acceptance records
317221 CFR 820.198(c)Investigation of device failures
317321 CFR 820.198(d)Evaluation, timeliness, identification
323321 CFR 820.72(a)Calibration, Inspection, etc. Procedures Lack of or Inadequ
324121 CFR 820.72(b)(1)Calibration standard traceability
326321 CFR 820.250(b)Sampling plans
326821 CFR 820.80(b)Incoming inspection, testing, verification
328321 CFR 820.90(a)Specific non-conforming product procedures
328621 CFR 820.90(b)(1)Procedures for product review,disposition lack of/inadequate
330221 CFR 820.100(a)(3)Identifying corrective & preventive actions
330421 CFR 820.100(a)(5)Changes to correct/prevent quality problems
336921 CFR 820.198(a)(1)Uniform and timely processing
337921 CFR 820.198(e)(5)Nature and details of complaint
339621 CFR 820.80(d)(1)DMR required activities {see also 820.181}
367021 CFR 820.20(c)Management review participants
367721 CFR 820.30(g)Design validation - software validation not performed
367821 CFR 820.30(g)Design Validation - Risk analysis not performed/inadequate
369021 CFR 820.100(a)(3)Identification of actions needed
369621 CFR 820.100(b)Documentation
370221 CFR 820.250(a)Appropriateness
407021 CFR 820.30(g)Design validation - documentation
418921 CFR 820.198(a)General
45421 CFR 820.40(a)Document review, approval by designated individual
47921 CFR 820.50Purchasing controls, Lack of or inadequate procedures
54621 CFR 820.75(a)Lack of or inadequate process validation
63121 CFR 803.17(a)(1)Lack of System for Event Evaluations
73121 CFR 803.50(a)(1)Report of Death or Serious Injury
73221 CFR 803.50(a)(2)Individual Report of Malfunction