FDA Inspection 716018 — Optovue, Inc.

Optovue, Inc. — FEI 3005950902

The FDA completed an inspection of Optovue, Inc. on Apr 7, 2011 in Fremont, CA. The inspection was classified NAI (No Action Indicated — no objectionable conditions or practices were found during the inspection). 6 citations were issued during this inspection. Product area: Devices.

Inspection details

Classification
No Action Indicated (NAI)
End Date
Apr 7, 2011
Fiscal Year
2011
Product Type
Devices
Project Area
Postmarket Assurance: Devices
Location
Fremont, CA (United States)

Citations

IDCFRDescription
1471621 CFR 820.30(f)Design verification - output does not meet input requirement
313021 CFR 820.100(a)Lack of or inadequate procedures
316821 CFR 820.198(a)Complaints
317021 CFR 820.198(b)Review and evaluation for investigation
367721 CFR 820.30(g)Design validation - software validation not performed
367821 CFR 820.30(g)Design Validation - Risk analysis not performed/inadequate