OPTOVUE, INC. FDA Inspection Record
OPTOVUE, INC. is an FDA-registered facility (FEI 3005950902) located in San Jose, CA, US. RegDataLab tracks 6 FDA inspections and 1 compliance action at this facility.
What FDA found at this site
- Latest inspection
- Mar 20, 2017: No Action Indicated (NAI)
- Inspections on record
- 6 (2 OAI, 1 VAI, 3 NAI)
- Last OAI inspection
- Sep 14, 2012
- Warning letters
- 1 listed by FDA for this FEI, latest Jun 11, 2010
- Recalls as the recalling firm
- 4, latest Jul 27, 2017
Inspections and what FDA cited
| Date | Result | What FDA cited |
|---|---|---|
| Mar 20, 2017 | No Action Indicated (NAI) | No citations listed |
| Apr 14, 2016 | No Action Indicated (NAI) | No citations listed |
| Nov 15, 2013 | Voluntary Action Indicated (VAI) | Lack of or inadequate complaint procedures (21 CFR 820.198(a)) Design control - no procedures (21 CFR 820.30(a)) Lack of or inadequate procedures (21 CFR 820.100(a)) |
| Sep 14, 2012 | Official Action Indicated (OAI) | Lack of or inadequate complaint procedures (21 CFR 820.198(a)) Design changes - Lack of or Inadequate Procedures (21 CFR 820.30(i)) |
| Apr 7, 2011 | No Action Indicated (NAI) | Design verification - output does not meet input requirement (21 CFR 820.30(f)) Lack of or inadequate procedures (21 CFR 820.100(a)) Complaints (21 CFR 820.198(a)) Review and evaluation for investigation (21 CFR 820.198(b)) and 2 more |
| Jan 8, 2010 | Official Action Indicated (OAI) | Training (21 CFR 820.25(b)) Design verification procedures (21 CFR 820.30(f)) Servicing - Lack of or inadequate procedures (21 CFR 820.200(a)) Processing MDRs {see also 803, 804} (21 CFR 820.198(a)(3)) and 3 more |
Recalls
| Date | Class | Recall | Product and reason |
|---|---|---|---|
| Jul 27, 2017 | Class II | Z-0008-2018 | iVue with Normative Database, Device Listing Number D185549.
Product Usage: The iVue… and 2 more products FDA determined that a 510(k) is needed for the Vault Mapping software (also known as Lens Fitting). |
| Mar 16, 2016 | Class II | Z-1498-2016 | iVue and iVue 500 (now called iScan) OCT systems with software version 2016.0.0.127 (USA… A software anomaly was detected that affects Optovue iVue and iVue 500 with Normative Database that may display images of the same eye… |
| May 14, 2013 | Class III | Z-1441-2013 | Optovue iVue with Normative Database (NDB) with Software Version 3.0 and 3.1.
Intended… Colors from the NDB comparison for Ganglion Cell Complex thickness in the singular OU report in iVue version 3.0 (US) and 3.1(international)… |
| Sep 7, 2012 | Class II | Z-0024-2013 | Reference materials "RETINAL OCT Analysis and Interpretation Method" shipped with iVue… Devices labeled for an intended use not included in the existing 510(k) or PMA. Specifically, "The Retina" poster provided with the systems… |
FDA record score
1 of 100. A count of this site's adverse public FDA records, weighted by how recent each one is. A higher number means more, or more recent, records. Each record below adds the points shown.
7 records behind this number
| Record | Date | Points x age weight | Adds |
|---|---|---|---|
| Class II recall Z-0008-2018, this FEI as the recalling firm | Jul 27, 2017 | 4 x 0.25 | 1 |
| Class II recall Z-1498-2016, this FEI as the recalling firm | Mar 16, 2016 | 10 years or older, so it adds no points | 0 |
| Inspection classified VAI | Nov 15, 2013 | 10 years or older, so it adds no points | 0 |
| Inspection classified OAI | Sep 14, 2012 | 10 years or older, so it adds no points | 0 |
| Class II recall Z-0024-2013, this FEI as the recalling firm | Sep 7, 2012 | 10 years or older, so it adds no points | 0 |
| Warning letter on FDA's dashboard, which does not show close-outs | Jun 11, 2010 | 10 years or older, so it adds no points | 0 |
| Inspection classified OAI | Jan 8, 2010 | 10 years or older, so it adds no points | 0 |
When each source was last refreshed
| Records | Last refreshed | Source as of | Status |
|---|---|---|---|
| Inspections | Nov 22, 2026 | Newest record Nov 9, 2026 | Current |
| Inspection citations | Nov 22, 2026 | Newest record Nov 5, 2026 | Current |
| Compliance actions (warning letters, injunctions and other action types) | Nov 22, 2026 | Newest record Nov 13, 2026 | Current |
| Warning letter text | Nov 22, 2026 | Newest record Nov 17, 2026 | Current |
| Recalls (Enforcement Report) | Nov 22, 2026 | Newest record Nov 18, 2026 | Current |
| Form 483s and inspection documents | Nov 18, 2026 | Newest record Nov 17, 2026 | Current |
| Import refusals | Nov 22, 2026 | Newest record Nov 12, 2026 | Current |
Reviewing OPTOVUE, INC. as a supplier? Request a free FDA snapshot: name up to five suppliers and get a PDF of their public FDA record, with sources and dates. You do not need an API connection. Each snapshot is prepared by hand.
Or check OPTOVUE, INC. yourself in Claude, Cursor or ChatGPT: get the full FDA risk summary, with inspections, warning letters, Form 483s and import alerts by site.
Facility details
- Company
- Optovue, Inc.
- Location
- San Jose, CA (US)
- Source
- device