OPTOVUE, INC. FDA Inspection Record

FEI 3005950902 · Inspections, Form 483s, warning letters and import refusals

OPTOVUE, INC. is an FDA-registered facility (FEI 3005950902) located in San Jose, CA, US. RegDataLab tracks 6 FDA inspections and 1 compliance action at this facility.

What FDA found at this site

Latest inspection
Mar 20, 2017: No Action Indicated (NAI)
Inspections on record
6 (2 OAI, 1 VAI, 3 NAI)
Last OAI inspection
Sep 14, 2012
Warning letters
1 listed by FDA for this FEI, latest Jun 11, 2010
Recalls as the recalling firm
4, latest Jul 27, 2017

A Form 483 lists an inspector's observations. OAI means FDA recommends regulatory and/or administrative action. A warning letter is FDA's formal enforcement notice. No record here does not establish that a site is safe. Check this site on FDA's Data Dashboard.

Inspections and what FDA cited

Newest first. Citations are the inspector observations FDA publishes for the inspection.

DateResultWhat FDA cited
Mar 20, 2017No Action Indicated (NAI)No citations listed
Apr 14, 2016No Action Indicated (NAI)No citations listed
Nov 15, 2013Voluntary Action Indicated (VAI)Lack of or inadequate complaint procedures (21 CFR 820.198(a))
Design control - no procedures (21 CFR 820.30(a))
Lack of or inadequate procedures (21 CFR 820.100(a))
Sep 14, 2012Official Action Indicated (OAI)Lack of or inadequate complaint procedures (21 CFR 820.198(a))
Design changes - Lack of or Inadequate Procedures (21 CFR 820.30(i))
Apr 7, 2011No Action Indicated (NAI)Design verification - output does not meet input requirement (21 CFR 820.30(f))
Lack of or inadequate procedures (21 CFR 820.100(a))
Complaints (21 CFR 820.198(a))
Review and evaluation for investigation (21 CFR 820.198(b))
and 2 more
Jan 8, 2010Official Action Indicated (OAI)Training (21 CFR 820.25(b))
Design verification procedures (21 CFR 820.30(f))
Servicing - Lack of or inadequate procedures (21 CFR 820.200(a))
Processing MDRs {see also 803, 804} (21 CFR 820.198(a)(3))
and 3 more

Recalls

Newest first. Recalls FDA's Enforcement Report lists with this FEI as the recalling firm. The recalling firm is not always the site that made the product. Search FDA's Enforcement Report by recall number.

DateClassRecallProduct and reason
Jul 27, 2017Class IIZ-0008-2018iVue with Normative Database, Device Listing Number D185549. Product Usage: The iVue… and 2 more products
FDA determined that a 510(k) is needed for the Vault Mapping software (also known as Lens Fitting).
Mar 16, 2016Class IIZ-1498-2016iVue and iVue 500 (now called iScan) OCT systems with software version 2016.0.0.127 (USA…
A software anomaly was detected that affects Optovue iVue and iVue 500 with Normative Database that may display images of the same eye…
May 14, 2013Class IIIZ-1441-2013Optovue iVue with Normative Database (NDB) with Software Version 3.0 and 3.1. Intended…
Colors from the NDB comparison for Ganglion Cell Complex thickness in the singular OU report in iVue version 3.0 (US) and 3.1(international)…
Sep 7, 2012Class IIZ-0024-2013Reference materials "RETINAL OCT Analysis and Interpretation Method" shipped with iVue…
Devices labeled for an intended use not included in the existing 510(k) or PMA. Specifically, "The Retina" poster provided with the systems…

FDA record score

1 of 100. A count of this site's adverse public FDA records, weighted by how recent each one is. A higher number means more, or more recent, records. Each record below adds the points shown.

Counted on Nov 22, 2026, method version 1. It counts records and does not judge the site: a low number does not establish that a site is safe, and sites that FDA inspects more often can collect more points. How this is counted.

7 records behind this number
RecordDatePoints x age weightAdds
Class II recall Z-0008-2018, this FEI as the recalling firmJul 27, 20174 x 0.251
Class II recall Z-1498-2016, this FEI as the recalling firmMar 16, 201610 years or older, so it adds no points0
Inspection classified VAINov 15, 201310 years or older, so it adds no points0
Inspection classified OAISep 14, 201210 years or older, so it adds no points0
Class II recall Z-0024-2013, this FEI as the recalling firmSep 7, 201210 years or older, so it adds no points0
Warning letter on FDA's dashboard, which does not show close-outsJun 11, 201010 years or older, so it adds no points0
Inspection classified OAIJan 8, 201010 years or older, so it adds no points0

When each source was last refreshed

FDA sources behind this page and when RegDataLab last refreshed each one
RecordsLast refreshedSource as ofStatus
InspectionsNov 22, 2026Newest record Nov 9, 2026Current
Inspection citationsNov 22, 2026Newest record Nov 5, 2026Current
Compliance actions (warning letters, injunctions and other action types)Nov 22, 2026Newest record Nov 13, 2026Current
Warning letter textNov 22, 2026Newest record Nov 17, 2026Current
Recalls (Enforcement Report)Nov 22, 2026Newest record Nov 18, 2026Current
Form 483s and inspection documentsNov 18, 2026Newest record Nov 17, 2026Current
Import refusalsNov 22, 2026Newest record Nov 12, 2026Current

Reviewing OPTOVUE, INC. as a supplier? Request a free FDA snapshot: name up to five suppliers and get a PDF of their public FDA record, with sources and dates. You do not need an API connection. Each snapshot is prepared by hand.

Check this supplier

Or check OPTOVUE, INC. yourself in Claude, Cursor or ChatGPT: get the full FDA risk summary, with inspections, warning letters, Form 483s and import alerts by site.

Get a free API key

Facility details

Company
Optovue, Inc.
Location
San Jose, CA (US)
Source
device